R&S FSW-K7 User Manual page 292

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Parameters:
<AverageCount>
Manual operation:
[SENSe:]AVERage<n>[:STATe<t>] <State>
This command turns averaging for a particular trace in a particular window on and off.
Suffix:
<n>
<t>
Parameters:
<State>
[SENSe:]AVERage<n>:TYPE <Mode>
This command selects the trace averaging mode.
Suffix:
<n>
Parameters:
<Mode>
Example:
Manual operation:
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion] <Detector>
Defines the trace detector to be used for trace analysis.
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
If you set an average count of 0 or 1, the application performs
one single sweep in single sweep mode.
In continuous sweep mode, if the average count is set to 0, a
moving average over 10 sweeps is performed.
Range:
0 to 200000
*RST:
0
See
"Sweep / Average Count"
See
"Average Count"
.
Window
Trace
ON | OFF | 1 | 0
.
Window
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
AVER:TYPE LIN
Switches to linear average calculation.
See
"Average Mode"
Configuring the Measurement
on page 109
on page 136
on page 135
292