R&S FSW-K7 User Manual page 201

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
[SENSe:]CORRection:CVL:DATA <Freq>,<Level>
This command defines the reference values of the selected conversion loss tables.
The values are entered as a set of frequency/level pairs. A maximum of 50 frequency/
level pairs may be entered. Before this command can be performed, the conversion
loss table must be selected (see
on page 202).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Freq>
<Level>
Example:
Manual operation:
[SENSe:]CORRection:CVL:HARMonic <HarmOrder>
This command defines the harmonic order for which the conversion loss table is to be
used. This setting is checked against the current mixer setting before the table can be
assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<HarmOrder>
Example:
Manual operation:
[SENSe:]CORRection:CVL:MIXer <Type>
This command defines the mixer name in the conversion loss table. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
numeric value
The frequencies have to be sent in ascending order.
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:DATA 1MHZ,-30DB,2MHZ,-40DB
See
"Position/Value"
[SENSe:]CORRection:CVL:SELect
numeric value
Range:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:HARM 3
See
"Harmonic Order"
[SENSe:]CORRection:CVL:SELect
[SENSe:]CORRection:CVL:SELect
on page 65
on page 202.
2 to 65
on page 64
on page 202).
Configuring the Measurement
201