R&S FSW-K7 User Manual page 258

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
TRIGger[:SEQuence]:IFPower:HOLDoff <Period>
This command defines the holding time before the next trigger event.
Note that this command can be used for any trigger source, not just IF Power
(despite the legacy keyword).
Note: If you perform gated measurements in combination with the IF Power trigger, the
R&S FSW ignores the holding time for frequency sweep, FFT sweep, zero span and
I/Q data measurements.
Parameters:
<Period>
Example:
Manual operation:
TRIGger[:SEQuence]:IFPower:HYSTeresis <Hysteresis>
This command defines the trigger hysteresis, which is only available for "IF Power" trig-
ger sources.
Parameters:
<Hysteresis>
Example:
Manual operation:
TRIGger[:SEQuence]:LEVel:BBPower <Level>
This command sets the level of the baseband power trigger.
This command is available for the optional Digital Baseband Interface and the optional
Analog Baseband Interface.
Parameters:
<Level>
Example:
Manual operation:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
Range:
0 s to 10 s
*RST:
0 s
TRIG:SOUR EXT
Sets an external trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
See
"Trigger Holdoff"
Range:
3 dB to 50 dB
*RST:
3 dB
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
See
"Hysteresis"
on page 102
Range:
-50 dBm to +20 dBm
*RST:
-20 dBm
TRIG:LEV:BBP -30DBM
See
"Trigger Level"
on page 101
Configuring the Measurement
on page 103
258

Advertisement

Table of Contents
loading

Table of Contents