For more information, see the R&S FSW I/Q Analyzer and I/Q Input User Manual.
Remote command:
CALibration:AIQ:HATiming[:STATe]
Center Frequency
Defines the center frequency for analog baseband input.
For real-type baseband input (I or Q only), the center frequency is always 0 Hz.
Note: If the analysis bandwidth to either side of the defined center frequency exceeds
the minimum frequency (0 Hz) or the maximum frequency (40 MHz/80 MHz), an error
is displayed. In this case, adjust the center frequency or the analysis bandwidth.
Remote command:
[SENSe:]FREQuency:CENTer
5.3.1.5

Probe Settings

Access: INPUT / OUTPUT > "Input Source Config" > "Probes"
For each possible probe connector (Baseband Input I, Baseband Input Q), the detec-
ted type of probe, if any, is displayed. The following information is provided for each
connected probe:
Probe name
Serial number
Rohde & Schwarz part number
Type of probe ("Differential", "Single Ended")
For more information on using probes with an R&S FSW, see the R&S FSW User
Manual.
For general information on the R&S
Common Mode
Microbutton
User Manual 1173.9240.02 ─ 23
on page 242
Offset................................................................................................... 71
Action........................................................................................................ 71
on page 211
®
RTO probes, see the device manuals.
Configuration
Input and Frontend Settings
70