R&S FSW-K7 User Manual page 95

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Remote command:
INPut:IQ:FULLscale:AUTO
INPut:IQ:FULLscale[:LEVel]
5.3.4 Frequency
Access: "Overview" > "Input/ Frontend" > "Frequency" tab
Center
Center Frequency
Center frequency
Defines the center frequency of the signal in Hertz.
The allowed range of values for the center frequency depends on the frequency span.
span > 0: span
f
and span
max
Remote command:
[SENSe:]FREQuency:CENTer
Center Frequency Stepsize
Defines the step size of the center frequency. The step size can be coupled to the
demodulation bandwidth, or it can be manually set to a fixed value.
"0.1 * Demod
BW"
"0.5 * Demod
BW"
User Manual 1173.9240.02 ─ 23
frequency...........................................................................................................95
Stepsize...........................................................................................95
/2 ≤ f
≤ f
min
center
max
depend on the instrument and are specified in the data sheet.
min
Sets the step size for the center frequency to 10 % of the demodula-
tion bandwidth.
This is the default setting.
Sets the step size for the center frequency to 50 % of the demodula-
tion bandwidth.
on page 210
on page 210
– span
/2
min
on page 242
Configuration
Input and Frontend Settings
95

Advertisement

Table of Contents
loading

Table of Contents