R&S
FSW-K7
11.4.8.3
Configuring the Demodulation Spectrum
The demodulation spectrum defines which span of the demodulated data is evaluated.
●
AF
●
RF
AF evaluation
These settings are only available for AF Spectrum evaluations, not in the time domain.
[SENSe:]ADEMod<n>:AF:CENTer
[SENSe:]ADEMod<n>:AF:SPAN
[SENSe:]ADEMod<n>:AF:SPAN:FULL
[SENSe:]ADEMod<n>:AF:STARt
[SENSe:]ADEMod<n>:AF:STOP
[SENSe:]ADEMod<n>:AF:CENTer <Frequency>
This command sets the center frequency for AF spectrum result display.
Suffix:
<n>
Parameters:
<Frequency>
Manual operation:
[SENSe:]ADEMod<n>:AF:SPAN <Span>
This command sets the span (around the center frequency) for AF spectrum result dis-
play.
The span is limited to DBW/2 (see
Suffix:
<n>
Parameters:
<Span>
Example:
Manual operation:
[SENSe:]ADEMod<n>:AF:SPAN:FULL
This command sets the maximum span for AF spectrum result display.
The maximum span corresponds to DBW/2 (see
on page 254).
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
evaluation........................................................................................................ 271
evaluation........................................................................................................ 272
......................................................................................271
.....................................................................................272
......................................................................................272
.
irrelevant
*RST:
See
"AF Center"
.
irrelevant
*RST:
ADEM:AF:SPAN 200 kHz
Sets the AF span to 200 kHz
See
"AF Span"
...................................................................................271
.............................................................................271
1.25 MHz
on page 114
[SENSe:]BWIDth:DEMod
9 MHz
on page 114
[SENSe:]BWIDth:DEMod
Configuring the Measurement
on page 254).
271