R&S FSW-K7 User Manual page 67

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Full Scale Level
The "Full Scale Level" defines the level and unit that should correspond to an I/Q sam-
ple with the magnitude "1".
If "Auto" is selected, the level is automatically set to the value provided by the connec-
ted device.
Remote command:
INPut:DIQ:RANGe[:UPPer]
INPut:DIQ:RANGe[:UPPer]:UNIT
INPut:DIQ:RANGe[:UPPer]:AUTO
Adjust Reference Level to Full Scale Level
If enabled, the reference level is adjusted to the full scale level automatically if any
change occurs.
Remote command:
INPut:DIQ:RANGe:COUPling
Connected Instrument
Displays the status of the Digital Baseband Interface connection.
If an instrument is connected, the following information is displayed:
Name and serial number of the instrument connected to the Digital Baseband Inter-
face
Used port
Sample rate of the data currently being transferred via the Digital Baseband Inter-
face
Level and unit that corresponds to an I/Q sample with the magnitude "1"
Level), if provided by connected instrument
Remote command:
INPut:DIQ:CDEVice
DigIConf
Starts the optional R&S DigIConf application. This function is available in the In-/Output
menu, but only if the optional software is installed.
Note that R&S DigIConf requires a USB connection (not LAN!) from the
R&S FSW to the R&S EX-IQ-BOX in addition to the Digital Baseband Interface
connection. R&S DigIConf version 2.20.360.86 Build 170 or higher is required.
To return to the R&S FSW application, press any key. The R&S FSW application is dis-
played with the "Input/Output" menu, regardless of which key was pressed.
For details on the R&S DigIConf application, see the "R&S
face Module R&S
Note: If you close the R&S DigIConf window using the "Close" icon, the window is
minimized, not closed.
If you select the "File > Exit" menu item in the R&S DigIConf window, the application is
closed. Note that in this case the settings are lost and the R&S EX-IQ-BOX functional-
ity is no longer available until you restart the application using the "DigIConf" softkey in
the R&S FSW once again.
User Manual 1173.9240.02 ─ 23
on page 214
on page 214
on page 214
on page 214
on page 213
®
DigIConf Software Operating Manual".
Configuration
Input and Frontend Settings
(Full Scale
®
EX-IQ-BOX Digital Inter-
67