R&S FSW-K7 User Manual page 276

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
[SENSe:]FILTer<n>:DEMPhasis[:STATe] <State>
This command activates/deactivates the selected deemphasis for the specified evalua-
tion.
For details about deemphasis refer to
Suffix:
<n>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]FILTer<n>:HPASs:FREQuency[:ABSolute] <FilterType>
This command selects the high pass filter type for the specified evaluation.
For details on the high pass filters refer to
Suffix:
<n>
Parameters:
<FilterType>
Example:
Manual operation:
[SENSe:]FILTer<n>:HPASs:FREQuency:MANual <Frequency>
This command selects the cutoff frequency of the high pass filter for the specified eval-
uation.
For details on the high pass filters refer to
Suffix:
<n>
Parameters:
<Frequency>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
"Deemphasis"
.
Window
ON | OFF | 1 | 0
*RST:
0
FILT:DEMP ON
Activates the selected deemphasis.
See
"Deemphasis"
on page 119
"High Pass"
.
Window
20 Hz | 50 Hz | 300 Hz
*RST:
300Hz
Default unit: Hz
FILT:HPAS:FREQ 300Hz
Selects the high pass filter for the demodulation bandwidth
range from 800 Hz to 8 MHz.
See
"High Pass"
on page 117
"High Pass"
.
Window
numeric value
Range:
0 to 3 MHz
*RST:
15kHz
Configuring the Measurement
on page 119.
on page 117.
on page 117.
276