R&S FSW-K7 User Manual page 268

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
[SENSe:]ADEMod<n>:PM:RPOint[:X] <Time>
This command determines the position where the phase of the PM-demodulated signal
is set to 0 rad. The maximum possible value depends on the measurement time
selected in the instrument; this value is output in response to the query
ADEM:PM:RPO:X? MAX.
Suffix:
<n>
Parameters:
<Time>
Example:
Manual operation:
[SENSe:]ADEMod<n>:SQUelch[:STATe] <State>
This command activates the squelch function, i.e. if the signal falls below a defined
threshold (see
lated data is automatically set to 0.
Suffix:
<n>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]ADEMod<n>:SQUelch:LEVel <Threshold>
This command defines the level threshold below which the demodulated data is set to
0 if squelching is enabled (see
on page 268).
Suffix:
<n>
Parameters:
<Threshold>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
.
irrelevant
0 s to measurement time
*RST:
0 s
ADEM:PM:RPO 500us
Sets the position where the phase to 0 rad setting to 500 µs.
See
"Zero Phase Reference Position (PM Time Domain only)"
on page 113
[SENSe:]ADEMod<n>:SQUelch:LEVel
.
irrelevant
ON | OFF | 1 | 0
*RST:
0
DEM:SQU ON
Signals below the level threshold are squelched.
See
"Squelch State"
[SENSe:]ADEMod<n>:SQUelch[:STATe]
.
irrelevant
numeric value
The absolute threshold level
Range:
-150 dBm to 30 dBm
*RST:
-40 dBm
Configuring the Measurement
on page 268), the demodu-
on page 111