Trace / Data Export Configuration - R&S FSW-K7 User Manual

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Remote command:
[SENSe:]AVERage<n>:TYPE
Average Count
Determines the number of averaging or maximum search procedures If the trace
modes "Average", "Max Hold" or "Min Hold" are set.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, Max Hold or Min Hold operations are
performed.
Remote command:
[SENSe:]AVERage<n>:COUNt
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
Function
Preset All Traces
Set Trace Mode
Max | Avg | Min
Set Trace Mode
Max | ClrWrite | Min
Trace 1/Trace 2/Trace 3/Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]

6.2 Trace / Data Export Configuration

Access: "Save" > "Export" > "(Trace) Export Config"
Or: TRACE > "Trace Config" > "Trace/Data Export"
The standard data management functions (e.g. saving or loading instrument settings)
that are available for all R&S FSW applications are not described here.
See the R&S FSW User Manual for a description of the standard functions.
User Manual 1173.9240.02 ─ 23
on page 292
on page 291
Trace Settings
Trace 1:
Clear Write
Traces 2-6:
Blank
Trace 1:
Max Hold
Trace 2:
Average
Trace 3:
Min Hold
Traces 4-6:
Blank
Trace 1:
Max Hold
Trace 2:
Clear Write
Trace 3:
Min Hold
Traces 4-6:
Blank

Trace / Data Export Configuration

on page 289
Analysis
136

Advertisement

Table of Contents
loading

Table of Contents