R&S FSW-K7 User Manual page 272

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Suffix:
<n>
Example:
Manual operation:
[SENSe:]ADEMod<n>:AF:STARt <Frequency>
This command sets the start frequency for AF spectrum result display.
Suffix:
<n>
Parameters:
<Frequency>
Example:
Manual operation:
[SENSe:]ADEMod<n>:AF:STOP <Frequency>
This command sets the stop frequency for AF spectrum result display.
Suffix:
<n>
Parameters:
<Frequency>
Example:
Manual operation:
RF evaluation
These settings are only available for RF evaluation, both in time and frequency
domain.
Useful commands described elsewhere
[SENSe:]FREQuency:CENTer
[SENSe:]BWIDth:DEMod
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
.
irrelevant
ADEM:BAND 5 MHz
Sets the demodulation bandwidth to 5 MHz
ADEM:AF:SPAN:FULL
Sets the AF span to 2.5 MHz
See
"AF Full Span"
on page 114
.
irrelevant
*RST:
0 MHz
ADEM:AF:STAR 0 kHz
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500 kHz
Sets the AF stop frequency to 500 kHz
See
"AF Start"
on page 114
.
irrelevant
*RST:
9 MHz
ADEM:AF:STAR 0 kHz
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500 kHz
Sets the AF stop frequency to 500 kHz
See
"AF Stop"
on page 114
on page 242
on page 254
Configuring the Measurement
272

Advertisement

Table of Contents
loading

Table of Contents