SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe <Rate>
Determines whether the 10 GHz mode (default) or 20 GHz mode of the connected
oscilloscope is used. The 20 GHZ mode achieves a higher decimation gain, but
reduces the record length by half.
Return values:
<Rate>
Example:
Manual operation:
SYSTem:COMMunicate:RDEVice:OSCilloscope:TCPip <Address>
Defines the TCPIP address or computer name of the oscilloscope connected to the
R&S FSW via LAN.
Note: The IP address is maintained after a PRESET, and is transferred between appli-
cations.
Parameters:
<Address>
Example:
Example:
Manual operation:
SYSTem:COMMunicate:RDEVice:OSCilloscope:VDEVice?
Queries whether the connected instrument is supported by the 2 GHz bandwidth
extension option(R&S FSW-B2000).
For details see the 2 GHz bandwidth extension basics chapter in the R&S FSW I/Q
Analyzer and I/Q Input User Manual.
Return values:
<State>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
10 GHz | 20 GHz
No other sample rate values are allowed.
*RST:
10 GHz
TRAC:IQ:SRAT?
//Result: 100000000
TRAC:IQ:RLEN?
//Result: 3128
SYST:COMM:RDEV:OSC:SRAT 20GHZ
TRAC:IQ:SRAT?
//Result: 200000000
TRAC:IQ:RLEN?
//Result: 1564
See
"Oscilloscope Sample Rate"
computer name or IP address
SYST:COMM:RDEV:OSC:TCP '192.0.2.0'
SYST:COMM:RDEV:OSC:TCP 'FSW43-12345'
See
"TCPIP Address or Computer name"
ON | 1
Instrument is supported
OFF | 0
Instrument is not supported