R&S FSW-K7 User Manual page 223

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Example:
Manual operation:
[SENSe:]PMETer<p>:DCYCle:VALue <Percentage>
This command defines the duty cycle for the correction of pulse signals.
The power sensor uses the duty cycle in combination with the mean power to calculate
the power of the pulse.
Suffix:
<p>
Parameters:
<Percentage>
Example:
Manual operation:
[SENSe:]PMETer<p>:FREQuency <Frequency>
This command defines the frequency of the power sensor.
Suffix:
<p>
Parameters:
<Frequency>
Example:
Manual operation:
[SENSe:]PMETer<p>:FREQuency:LINK <Coupling>
This command selects the frequency coupling for power sensor measurements.
Suffix:
<p>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
PMET2:DCYC:STAT ON
See
"Duty Cycle"
on page 86
.
1...4
Power sensor
Range:
0.001 to 99.999
*RST:
99.999
Default unit: %
PMET2:DCYC:STAT ON
Activates the duty cycle correction.
PMET2:DCYC:VAL 0.5
Sets the correction value to 0.5%.
See
"Duty Cycle"
on page 86
.
1...4
Power sensor index
The available value range is specified in the data sheet of the
power sensor in use.
*RST:
50 MHz
PMET2:FREQ 1GHZ
Sets the frequency of the power sensor to 1 GHz.
See
"Frequency Manual"
.
1...4
Power sensor index
Configuring the Measurement
on page 85
223