R&S FSW-K7 User Manual page 228

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Manual operation:
[SENSe:]PMETer<p>:TRIGger:HYSTeresis <Hysteresis>
This command defines the trigger hysteresis for external power triggers.
The hysteresis in dB is the value the input signal must stay below the IF power trigger
level in order to allow a trigger to start the measurement.
Suffix:
<p>
Parameters:
<Hysteresis>
Example:
Manual operation:
[SENSe:]PMETer<p>:TRIGger:LEVel <Level>
This command defines the trigger level for external power triggers.
Suffix:
<p>
Parameters:
<Level>
Example:
Manual operation:
[SENSe:]PMETer<p>:TRIGger:SLOPe <Edge>
This command selects the trigger condition for external power triggers.
Suffix:
<p>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
See
"Trigger Holdoff"
.
1...4
Power sensor index
Range:
3 dB to 50 dB
Increment: 1 dB
*RST:
0 dB
PMET2:TRIG:HYST 10
Sets the hysteresis of the trigger to 10 dB.
See
"Hysteresis"
on page 87
.
1...4
Power sensor index
-20 to +20 dBm
Range:
-20 dBm to 20 dBm
*RST:
-10 dBm
PMET2:TRIG:LEV -10 dBm
Sets the level of the trigger
See
"External Trigger Level"
.
1...4
Power sensor index
Configuring the Measurement
on page 87
on page 87
228