R&S FSW-K7 User Manual page 291

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Parameters:
<TraceMode>
Example:
[SENSe:]AVERage<n>:COUNt <AverageCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
WRITe
Overwrite mode: the trace is overwritten by each sweep. This is
the default setting.
AVERage
The average is formed over several sweeps.
The
Sweep / Average Count
ing procedures.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S FSW saves the sweep result in the trace mem-
ory only if the new value is greater than the previous one.
MINHold
The minimum value is determined from several measurements
and displayed. The R&S FSW saves the sweep result in the
trace memory only if the new value is lower than the previous
one.
VIEW
The current contents of the trace memory are frozen and dis-
played.
OFF
Hides the selected trace.
*RST:
WRITe,OFF,OFF,OFF,OFF,OFF
ADEM:AM AVER,MAXH,MINH,OFF,OFF,OFF
Determines average, max hold and min hold values simultane-
ously for the traces 1-3 of the RF time domain evaluation.
ADEM:AM WRIT,OFF,OFF,OFF,OFF,OFF
Determines only the current measurement values for trace 1.
ADEM:AM OFF,OFF,OFF,OFF,OFF,OFF
Switches AM demodulation off.
.
irrelevant
Configuring the Measurement
determines the number of averag-
291

Advertisement

Table of Contents
loading

Table of Contents