R&S FSW-K7 User Manual page 298

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
INITiate<n>:SEQuencer:MODE <Mode>
This command selects the way the R&S FSW application performs measurements
sequentially.
Before this command can be executed, the Sequencer must be activated (see
SYSTem:SEQuencer
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S FSW User Manual.
Note: In order to synchronize to the end of a sequential measurement using *OPC,
*OPC? or *WAI you must use SINGle Sequence mode.
For details on synchronization see the "Remote Basics" chapter in the R&S FSW User
Manual.
Suffix:
<n>
Parameters:
<Mode>
Example:
INITiate<n>:SEQuencer:REFResh[:ALL]
This function is only available if the Sequencer is deactivated
SYST:SEQ:OFF) and only in MSRA / MSRT mode.
The data in the capture buffer is re-evaluated by all active MSRA / MSRT slave appli-
cations.
Suffix:
<n>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
on page 299).
.
irrelevant
SINGle
Each measurement is performed once (regardless of the chan-
nel's sweep mode), considering each channels' sweep count,
until all measurements in all active channels have been per-
formed.
CONTinuous
The measurements in each active channel are performed one
after the other, repeatedly (regardless of the channel's sweep
mode), in the same order, until the Sequencer is stopped.
CDEFined
First, a single sequence is performed. Then, only those channels
in continuous sweep mode (INIT:CONT ON) are repeated.
*RST:
CONTinuous
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
.
irrelevant
Capturing Data and Performing Sweeps
(SYSTem:SEQuencer
298