R&S FSW-K7 User Manual page 373

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Parameters:
<State>
Manual operation:
CALCulate<n>:RTMS:ALINe[:VALue] <Position>
This command defines the position of the analysis line for all time-based windows in all
MSRT slave applications and the MSRT Master.
Suffix:
<n>
Parameters:
<Position>
Manual operation:
CALCulate<n>:RTMS:WINDow<n>:IVAL?
This command queries the analysis interval for the window specified by the WINDow
suffix <n> (the CALC suffix is irrelevant). This command is only available in application
measurement channels, not the MSRT View or MSRT Master.
Suffix:
<n>
Return values:
<IntStart>
<IntStop>
Usage:
[SENSe:]RTMS:CAPTure:OFFSet <Offset>
This setting is only available for slave applications in MSRT mode, not for the MSRT
Master. It has a similar effect as the trigger offset in other measurements.
Parameters:
<Offset>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
ON | OFF | 1 | 0
*RST:
1
See
"Show Line"
on page 161
.
irrelevant
Position of the analysis line in seconds. The position must lie
within the measurement time (pretrigger + posttrigger) of the
MSRT measurement.
Default unit: s
See
"Position"
on page 161
.
Window
Start value of the analysis interval in seconds
Default unit: s
Stop value of the analysis interval in seconds
Query only
This parameter defines the time offset between the capture buf-
fer start and the start of the extracted slave application data. The
offset must be a positive value, as the slave application can only
analyze data that is contained in the capture buffer.
Range:
- [pretrigger time] to min (posttrigger time; sweep
time)
*RST:
0
Analyzing Results
373

Advertisement

Table of Contents
loading

Table of Contents