Measuring Phase Noise - R&S FSW-K7 User Manual

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:NEXT
CALCulate<n>:MARKer<m>:MAXimum:RIGHt
CALCulate<n>:MARKer<m>:MAXimum:LEFT
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT
Search Minimum
Sets the selected marker/delta marker to the minimum of the trace. If no marker is
active, marker 1 is activated.
Remote command:
CALCulate<n>:MARKer<m>:MINimum[:PEAK]
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK]
Search Next Minimum
Sets the selected marker/delta marker to the next (higher) minimum of the selected
trace. If no marker is active, marker 1 is activated.
Remote command:
CALCulate<n>:MARKer<m>:MINimum:NEXT
CALCulate<n>:MARKer<m>:MINimum:LEFT
CALCulate<n>:MARKer<m>:MINimum:RIGHt
CALCulate<n>:DELTamarker<m>:MINimum:NEXT
CALCulate<n>:DELTamarker<m>:MINimum:LEFT
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt

6.3.3 Measuring Phase Noise

Access: "Overview" > "Analysis" > "Marker Functions" > "Phase Noise"
or: MKR FUNC > "Select Marker Function" > "Phase Noise"
Phase noise is unintentional modulation of a carrier; it creates frequencies next to the
carrier frequency. A phase noise measurement consists of noise density measure-
ments at defined offsets from the carrier; the results are given in relation to the carrier
level (dBc).
In the Analog Demodulation application, phase noise measurement markers are avail-
able for the AF Spectrum result displays. For the FM Spectrum and PM Spectrum
result displays, the phase deviation in rad equals the phase noise at the marker posi-
tion. For AM Spectrum displays the marker result equals the amplitude noise at the
marker position.
The noise power density is measured at each marker for which the phase noise func-
tion is activated, and set in relation to the measured carrier power. A reference marker
is not required. In the marker table display, the phase noise is indicated as the marker
function result.
User Manual 1173.9240.02 ─ 23
Working with Markers in the R&S FSW Analog Demodulation application
on page 339
on page 340
on page 339
on page 342
on page 342
on page 341
on page 341
on page 343
on page 340
on page 340
on page 341
on page 343
on page 342
on page 343
Analysis
145