R&S FSW-K7 User Manual page 277

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Example:
Manual operation:
[SENSe:]FILTer<n>:HPASs[:STATe] <State>
This command activates/deactivates the selected high pass filter for the specified eval-
uation.
For details on the high pass filter refer to
Suffix:
<n>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]FILTer<n>:LPASs:FREQuency[:ABSolute] <FilterType>
This command selects the absolute low pass filter type for the specified evaluation
For details on the low pass filter refer to
Suffix:
<n>
Parameters:
<FilterType>
Example:
Manual operation:
[SENSe:]FILTer<n>:LPASs:FREQuency:MANual <Frequency>
This command selects the cutoff frequency of the low pass filter for the specified evalu-
ation.
For details on the low pass filter refer to
Suffix:
<n>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
FILT:HPAS:FREQ:MAN 3MHz
The AF results are restricted to frequencies lower than 3 MHz.
See
"High Pass"
on page 117
"High Pass"
.
Window
ON | OFF | 1 | 0
*RST:
0
FILT:HPAS ON
Activates the selected high pass filter.
See
"High Pass"
on page 117
"Low Pass"
.
Window
3kHz | 15kHz | 150kHz
*RST:
15kHz
FILT:LPAS:FREQ 150kHz
Selects the low pass filter for the demodulation bandwidth range
from 400 kHz to 16 MHz.
See
"Low Pass"
on page 117
"Low Pass"
.
Window
Configuring the Measurement
on page 117.
on page 117.
on page 117.
277