R&S FSW-K7 User Manual page 358

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Parameters:
<LimitLinePoints>
Manual operation:
CALCulate<n>:LIMit<k>:LOWer:MARGin <Margin>
This command defines an area around a lower limit line where limit check violations
are still tolerated.
Suffix:
<n>
<k>
Parameters:
<Margin>
Manual operation:
CALCulate<n>:LIMit<k>:LOWer:MODE <Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
<k>
Parameters:
<Mode>
Manual operation:
CALCulate<n>:LIMit<k>:LOWer:OFFSet <Offset>
This command defines an offset for a complete lower limit line.
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with
LIMit<k>:CONTrol[:DATA]. If not, the R&S FSW either adds
missing values or ignores surplus values.
The unit depends on
on page 360.
*RST:
Limit line state is OFF
See
"Data points"
on page 159
.
irrelevant
Limit line
numeric value
*RST:
0
Default unit: dB
See
"Margin"
on page 159
.
Window
Limit line
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
See
"X-Axis"
on page 159
CALCulate<n>:
CALCulate<n>:LIMit<k>:UNIT
Analyzing Results
358

Advertisement

Table of Contents
loading

Table of Contents