R&S FSW-K7 User Manual page 297

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Example:
Usage:
Manual operation:
INITiate<n>:SEQuencer:ABORt
This command stops the currently active sequence of measurements. The Sequencer
itself is not deactivated, so you can start a new sequence immediately using
INITiate<n>:SEQuencer:IMMediate
To deactivate the Sequencer use
Suffix:
<n>
Usage:
INITiate<n>:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer.
Its effect is similar to the
measurement.
Before this command can be executed, the Sequencer must be activated (see
SYSTem:SEQuencer
Suffix:
<n>
Example:
Usage:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
SYST:SEQ:OFF
Deactivates the scheduler
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a new data measurement and waits for the end of the
sweep.
INST:SEL 'IQ ANALYZER'
Selects the IQ Analyzer channel.
INIT:REFR
Refreshes the display for the I/Q Analyzer channel.
Event
See
"Refresh ( MSRA / MSRT only)"
on page 297.
SYSTem:SEQuencer
.
irrelevant
Event
INITiate<n>[:IMMediate]
on page 299).
.
irrelevant
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Event
Capturing Data and Performing Sweeps
on page 108
on page 299.
command used for a single
297

Advertisement