[SENSe:]FREQuency:CENTer:STEP:LINK <CouplingType>
This command couples and decouples the center frequency step size to the span or
the resolution bandwidth.
Parameters:
<CouplingType>
Example:
Manual operation:
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor <Factor>
This command defines a step size factor if the center frequency step size is coupled to
the span or the resolution bandwidth.
Parameters:
<Factor>
Example:
Manual operation:
11.4.5 Configuring the Vertical Axis (Amplitude, Scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
Amplitude
Configuring the
Configuring a
Scaling the
11.4.5.1
Amplitude Settings
Useful commands for amplitude configuration described elsewhere:
[SENSe:]ADJust:LEVel
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
SPAN
Couples the step size to the span. Available for measurements
in the frequency domain.
(for RF spectrum result display)
RBW
Couples the step size to the resolution bandwidth. Available for
measurements in the time domain.
(for all result displays except RF spectrum)
OFF
Decouples the step size.
*RST:
SPAN
FREQ:CENT:STEP:LINK SPAN
See
"Center Frequency Stepsize"
1 to 100 PCT
*RST:
10
FREQ:CENT:STEP:LINK:FACT 20PCT
See
"Center Frequency Stepsize"
Settings................................................................................................ 243
Attenuation...................................................................................245
Preamplifier.....................................................................................248
Y-Axis................................................................................................. 249
on page 286
Configuring the Measurement
on page 95
on page 95
243