R&S
FSW-K7
CALCulate<n>:LIMit<k>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on all limit lines in all measurement windows at the same time.
Suffix:
<n>, <k>
Example:
CALCulate<n>:LIMit<k>:FAIL?
This command queries the result of a limit check in the specified window.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
<k>
Return values:
<Result>
Example:
Usage:
11.8.2.4
Programming Example: Using Limit Lines
The following examples demonstrate how to work with limit lines in a remote environ-
ment.
Example: Configuring Limit
Example: Performing a Limit
Example: Configuring Limit Lines
This example demonstrates how to configure 2 limit lines - an upper and a lower limit -
for a measurement in a remote environment.
//------------- Configuing the limit lines ---------------------
CALC:LIM1:NAME 'FM1'
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
.
irrelevant
CALC:LIM:CLE
Deletes the result of the limit check.
INITiate<n>:CONTinuous
.
Window
Limit line
0
PASS
1
FAIL
INIT;*WAI
Starts a new sweep and waits for its end.
CALC2:LIM3:FAIL?
Queries the result of the check for limit line 3 in window 2.
Query only
Lines..........................................................................367
on page 295.
Check......................................................................369
Analyzing Results
367