R&S FSW-K7 User Manual page 85

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe]
on page 219
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
Zeroing Power Sensor
Starts zeroing of the power sensor.
For details on the zeroing process refer to the R&S FSW User Manual.
Remote command:
CALibration:PMETer<p>:ZERO:AUTO ONCE
Frequency Manual
Defines the frequency of the signal to be measured. The power sensor has a memory
with frequency-dependent correction factors. This allows extreme accuracy for signals
of a known frequency.
Remote command:
[SENSe:]PMETer<p>:FREQuency
Frequency Coupling
Selects the coupling option. The frequency can be coupled automatically to the center
frequency of the instrument or to the frequency of marker 1.
Remote command:
[SENSe:]PMETer<p>:FREQuency:LINK
Unit/Scale
Selects the unit with which the measured power is to be displayed. Available units are
dBm, dB, W and %.
If dB or % is selected, the display is relative to the reference value that is defined with
either the "Meas -> Ref" setting or the "Reference Value" setting.
Remote command:
UNIT<n>:PMETer<p>:POWer
UNIT<n>:PMETer<p>:POWer:RATio
Meas Time/Average
Selects the measurement time or switches to manual averaging mode. In general,
results are more precise with longer measurement times. The following settings are
recommended for different signal types to obtain stable and precise results:
"Short"
"Normal"
"Long"
"Manual"
User Manual 1173.9240.02 ─ 23
on page 223
on page 226
Stationary signals with high power (> -40dBm), because they require
only a short measurement time and short measurement time provides
the highest repetition rates.
Signals with lower power or modulated signals
Signals at the lower end of the measurement range (<-50 dBm) or
Signals with lower power to minimize the influence of noise
Manual averaging mode. The average count is set with the
Count (Number of Readings)
Input and Frontend Settings
on page 219
on page 220
on page 223
on page 226
setting.
Configuration
Average
85

Advertisement

Table of Contents
loading

Table of Contents