FSW-K7
Generator type
SMIQ03
SMIQ03B
SMIQ03E
SMIQ04B
SMIQ06B
SMJ03
SMJ06
SML01
SML02
SML03
1) Requires firmware version V2.10.x or higher on the signal generator
2) Requires firmware version V1.10.x or higher on the signal generator
3) Requires the option SMR-B11 on the signal generator
4) Requires firmware version V3.20.200 or higher on the signal generator
4.7.4.3
Generator Setup Files
For each signal generator type to be controlled by the R&S FSW a generator setup file
must be configured and stored on the R&S FSW. The setup file defines the frequency
and power ranges supported by the generator, as well as information required for com-
munication. For the signal generators listed in
Generators",
be edited or duplicated for varying measurement setups or other instruments.
The existing setup files can be displayed in an editor in read-only mode directly from
the "External Generator" configuration dialog box. From there, they can be edited and
stored under a different name, and are then available on the R&S FSW.
(For details see the R&S FSW User Manual).
4.7.4.4
Calibration Mechanism
A common measurement setup includes a signal generator, a device under test (DUT),
and a signal and spectrum analyzer. Therefore, it is useful to measure the attenuation
or gain caused by the cables and connectors from the signal generator and the signal
analyzer in advance. The known level offsets can then be removed from the measure-
ment results in order to obtain accurate information on the DUT.
Calculating the difference between the currently measured power and a reference
trace is referred to as calibration. Thus, the measurement results from the controlled
external generator - including the inherent distortions - can be used as a reference
trace to calibrate the measurement setup.
User Manual 1173.9240.02 ─ 23
TTL support
X
X
-
X
X
X
X
-
-
-
on page 35, default setup files are provided. If necessary, these files can
Receiving Data Input and Providing Data Output
Generator type
SMY01
SMY02
HP8254A
HP8257D
HP8340A
HP8648
HP ESG-A Series 1000A, 2000A,
3000A, 4000A
HP ESG B Series
Chapter 4.7.4.2, "Overview of Supported
Measurement Basics