Example:
Manual operation:
CALCulate<n>:DELTamarker<m>:TRACe <Trace>
This command selects the trace a delta marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
<m>
Parameters:
<Trace>
Example:
CALCulate<n>:DELTamarker<m>:X <Position>
This command moves a delta marker to a particular coordinate on the x-axis.
If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Suffix:
<m>
<n>
Parameters:
<Position>
Example:
Manual operation:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
CALC:DELT2 ON
Turns on delta marker 2.
See
"Marker State"
on page 139
See
"Marker Type"
on page 140
See
"Select Marker"
.
Window
Marker
Trace number the marker is assigned to.
CALC:DELT2:TRAC 2
Positions delta marker 2 on trace 2.
.
Marker
Window
Numeric value that defines the marker position on the x-axis.
The position is relative to the reference marker.
To select an absolute position you have to change the delta
marker mode with
CALCulate<n>:DELTamarker<m>:MODE
on page 323.
A query returns the absolute position of the delta marker.
Range:
The value range and unit depend on the measure-
ment and scale of the x-axis.
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
See
"Marker Position (X-value)"
on page 144
on page 140
Analyzing Results
325