Example:
Manual operation:
INPut:ATTenuation:AUTO <State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S FSW determines the signal level for optimal
internal data processing and sets the required attenuation accordingly.
This function is not available if the optional Digital Baseband Interface is active.
Parameters:
<State>
Example:
Manual operation:
INPut:ATTenuation:AUTO:MODE <OptMode>
Selects the priority for signal processing after the RF attenuation has been applied.
Parameters:
<OptMode>
Example:
INPut:EATT <Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (INP:EATT:AUTO OFF, see
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
This command requires the electronic attenuation hardware option.
It is not available if the optional Digital Baseband Interface is active.
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
See
"Attenuation Mode / Value"
ON | OFF | 0 | 1
*RST:
1
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
See
"Attenuation Mode / Value"
LNOise | LDIStortion
LNOise
Optimized for high sensitivity and low noise levels
LDIStortion
Optimized for low distortion by avoiding intermodulation
*RST:
LDIStortion
INP:ATT:AUTO:MODE LNO
INPut:EATT:AUTO
Configuring the Measurement
on page 90
on page 90
on page 247).