FSW-K7
INPut:DIQ:CDEVice
This command queries the current configuration and the status of the digital I/Q input
from the optional Digital Baseband Interface.
For details see the section "Interface Status Information" for the optional Digital Base-
band Interface in the R&S FSW I/Q Analyzer User Manual.
Return values:
<ConnState>
<DeviceName>
<SerialNumber>
<PortName>
<SampleRate>
<MaxTransferRate> Maximum data transfer rate of the connected device in Hz
<ConnProtState>
<PRBSTestState>
<SampleRateType> 0
<FullScaleLevel>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
Defines whether a device is connected or not.
0
No device is connected.
1
A device is connected.
Device ID of the connected device
Serial number of the connected device
Port name used by the connected device
Maximum or currently used sample rate of the connected device
in Hz (depends on the used connection protocol version; indica-
ted by <SampleRateType> parameter)
State of the connection protocol which is used to identify the
connected device.
Not Started
Has to be Started
Started
Passed
Failed
Done
State of the PRBS test.
Not Started
Has to be Started
Started
Passed
Failed
Done
Maximum sample rate is displayed
1
Current sample rate is displayed
The level (in dBm) that should correspond to an I/Q sample with
the magnitude "1" (if transferred from connected device);
If not available, 1.#QNAN (not a number) is returned
Configuring the Measurement
213