R&S FSW-K7 User Manual page 64

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Remote command:
[SENSe:]CORRection:CVL:SELect
Comment
An optional comment that describes the conversion loss table. The comment is user-
definable.
Remote command:
[SENSe:]CORRection:CVL:COMMent
Band
The waveguide or user-defined band to which the table applies. This setting is checked
against the current mixer setting before the table can be assigned to the range.
For a definition of the frequency range for the pre-defined bands, see
page
Remote command:
[SENSe:]CORRection:CVL:BAND
Harmonic Order
The harmonic order of the range to which the table applies. This setting is checked
against the current mixer setting before the table can be assigned to the range.
Remote command:
[SENSe:]CORRection:CVL:HARMonic
Bias
The bias current which is required to set the mixer to its optimum operating point. It
corresponds to the short-circuit current. The bias current can range from -10 mA to
10 mA. The actual bias current is lower because of the forward voltage of the mixer
diode(s).
Tip: You can also define the bias interactively while a preview of the trace with the
changed setting is displayed, see
Remote command:
[SENSe:]CORRection:CVL:BIAS
Mixer Name
Specifies the name of the external mixer to which the table applies. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Remote command:
[SENSe:]CORRection:CVL:MIXer
Mixer S/N
Specifies the serial number of the external mixer to which the table applies.
The specified number is checked against the currently connected mixer number before
the table can be assigned to the range.
Remote command:
[SENSe:]CORRection:CVL:SNUMber
User Manual 1173.9240.02 ─ 23
196.
on page 202
on page 200
on page 199
on page 201
"Bias Settings"
on page 60.
on page 199
on page 201
on page 202
Configuration
Input and Frontend Settings
table 11-3 on
64

Advertisement