R&S
FSW-K7
Figure 9-2: Default Analog Demodulation measurement result display
5. Set the measurement time (AQT) to 1 ms in order to measure 10 periods of the sig-
nal.
6. Adjust the y-axis scaling to the measured frequency deviation automatically by
selecting the "Scale Config" softkey and, in the "Scaling" tab, setting "AF Auto
Scale" to "ON".
Figure 9-3: Auto-scaled measurement of 10 signal periods (continuous)
7. Display the RF spectrum of the measured signal to determine the required demod-
ulation bandwidth. Select the "Display Config" softkey and add an "RF Spectrum"
window to the display.
User Manual 1173.9240.02 ─ 23
Measurement Example: Demodulating an FM Signal