R&S FSW-K7 User Manual page 38

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
The inherent frequency and power level distortions can be determined by connecting
the R&S FSW to the signal generator. The R&S FSW sends a predefined list of fre-
quencies to the signal generator (see also
cies",
at each frequency in the predefined list. The R&S FSW measures the signal and deter-
mines the level offsets to the expected values.
Saving calibration results
A reference dataset for the calibration results is stored internally as a table of value
pairs (frequency/level), one for each sweep point. The measured offsets can then be
used as calibration factors for subsequent measurement results.
The calibration can be performed using either transmission or reflection measure-
ments. The selected type of measurement used to determine the reference trace is
included in the reference dataset.
4.7.4.5
Normalization
Once the measurement setup has been calibrated and the reference trace is available,
subsequent measurement results can be corrected according to the calibration factors,
if necessary. This is done by subtracting the reference trace from the measurement
results. This process is referred to as normalization and can be activated or deactiva-
ted as required. If normalization is activated, "NOR" is displayed in the channel bar,
next to the indication that an external generator is being used ("Ext.Gen").The normal-
ized trace from the calibration sweep is a constant 0 dB line, as <calibration trace> -
<reference trace> = 0.
As long as the same settings are used for measurement as for calibration, the normal-
ized measurement results should not contain any inherent frequency or power distor-
tions. Thus, the measured DUT values are very accurate.
Approximate normalization
As soon as any of the calibration measurement settings are changed, the stored refer-
ence trace will no longer be identical to the new measurement results. However, if the
measurement settings do not deviate too much, the measurement results can still be
normalized approximately using the stored reference trace. This is indicated by the
"APX" label in the channel bar (instead of "NOR").
This is the case if one or more of the following values deviate from the calibration set-
tings:
Coupling (RBW, VBW, SWT)
Reference level, RF attenuation
Start or stop frequency
Output level of external generator
Detector (max. peak, min. peak, sample, etc.)
Frequency deviation at a maximum of 1001 points within the set sweep limits (cor-
responds to a doubling of the span)
User Manual 1173.9240.02 ─ 23
on page 41). The signal generator then sends a signal with the specified level
Receiving Data Input and Providing Data Output
Chapter 4.7.4.7, "Coupling the Frequen-
Measurement Basics
38

Advertisement

Table of Contents
loading

Table of Contents