R&S FSW-K7 User Manual page 330

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
CALCulate<n>:MARKer<m>:SGRam:FRAMe <Frame> | <Time>
CALCulate<n>:MARKer<m>:SPECtrogram:FRAMe <Frame> | <Time>
This command positions a marker on a particular frame.
Suffix:
<n>
<m>
Parameters:
<Frame>
<Time>
Example:
CALCulate<n>:MARKer<m>:SGRam:SARea <SearchArea>
CALCulate<n>:MARKer<m>:SPECtrogram:SARea <SearchArea>
This command defines the marker search area for all spectrogram markers in the mea-
surement channel.
Suffix:
<n>, <m>
Parameters:
<SearchArea>
CALCulate<n>:MARKer<m>:SGRam:XY:MAXimum[:PEAK]
CALCulate<n>:MARKer<m>:SPECtrogram:XY:MAXimum[:PEAK]
This command moves a marker to the highest level of the spectrogram.
Suffix:
<n>
<m>
Usage:
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
.
Window
Marker
Selects a frame directly by the frame number. Valid if the time
stamp is off.
The range depends on the history depth.
Selects a frame via its time stamp. Valid if the time stamp is on.
The number is the (negative) distance to frame 0 in seconds.
The range depends on the history depth.
CALC:MARK:SGR:FRAM -20
Sets the marker on the 20th frame before the present.
CALC:MARK2:SGR:FRAM -2s
Sets second marker on the frame 2 seconds ago.
.
irrelevant
VISible
Performs a search within the visible frames.
Note that the command does not work if the spectrogram is not
visible for any reason (e.g. if the display update is off).
MEMory
Performs a search within all frames in the memory.
*RST:
VISible
.
Window
Marker
Event
Analyzing Results
330

Advertisement

Table of Contents
loading

Table of Contents