R&S FSW-K7 User Manual page 135

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
"View"
"Blank"
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Detector
Defines the trace detector to be used for trace analysis.
"Auto"
"Type"
Remote command:
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
Hold
If activated, traces in "Min Hold", "Max Hold" and "Average" mode are not reset after
specific parameter changes have been made.
Normally, the measurement is started again after parameter changes, before the mea-
surement results are analyzed (e.g. using a marker). In all cases that require a new
measurement after parameter changes, the trace is reset automatically to avoid false
results (e.g. with span changes). For applications that require no reset after parameter
changes, the automatic reset can be switched off.
The default setting is off.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE:HCONtinuous
Average Mode
Defines the mode with which the trace is averaged over several sweeps. A different
averaging mode can be defined for each trace.
This setting is only applicable if trace mode "Average" is selected.
How many sweeps are averaged is defined by the
"Linear"
"Logarithmic"
"Power"
User Manual 1173.9240.02 ─ 23
The current contents of the trace memory are frozen and displayed.
Removes the selected trace from the display.
Selects the optimum detector for the selected trace and filter mode.
This is the default setting.
Defines the selected detector type.
Note: If the EMI (R&S FSW-K54) measurement option is installed
and the filter type "CISPR" is selected, additional detectors are availa-
ble, even if EMI measurement is not active.
The power level values are converted into linear units prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
For logarithmic scaling, the values are averaged in dBm. For linear
scaling, the behavior is the same as with linear averaging.
Activates linear power averaging.
The power level values are converted into unit Watt prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
Use this mode to average power values in Volts or Amperes cor-
rectly.
on page 287
on page 292
on page 293
on page 288
"Average Count"
Analysis
Trace Settings
on page 136.