R&S FSW-K7 User Manual page 296

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Parameters:
<State>
Example:
Manual operation:
INITiate<n>[:IMMediate]
This command starts a (single) new measurement.
With sweep count or average count > 0, this means a restart of the corresponding
number of measurements. With trace mode MAXHold, MINHold and AVERage, the
previous results are reset on restarting the measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
For details on synchronization see the "Remote Basics" chapter in the R&S FSW User
Manual.
Suffix:
<n>
Usage:
Manual operation:
INITiate<n>:REFResh
This function is only available if the Sequencer is deactivated
SYST:SEQ:OFF) and only for slave applications in MSRA / MSRT mode, not the
MSRA / MSRT Master.
The data in the capture buffer is re-evaluated by the currently active slave application
only. The results for any other slave applications remain unchanged.
The slave application channel must be selected before this command can be executed
(see
Suffix:
<n>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
INIT:CONT OFF
Switches the sweep mode to single sweep.
INIT:CONT ON
Switches the sweep mode to continuous sweep.
See
"Continuous Sweep/RUN CONT"
.
irrelevant
Event
See
"Single Sweep/ RUN SINGLE"
INSTrument[:SELect]
.
irrelevant
Capturing Data and Performing Sweeps
1
on page 186).
on page 107
on page 108
(SYSTem:SEQuencer
296