FSW-K7
CALCulate<n>:MARKer<m>:Y?
This command queries the position of a marker on the y-axis.
If necessary, the command activates the marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
If the analog demodulator (option Analog Demodulation, R&S FSW–K7) is activated,
the query result is output in the following units in the specified window:
Result display
AM
FM
PM
RF
Suffix:
<n>
<m>
Return values:
<Result>
Example:
Usage:
Manual operation:
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off all delta markers.
Suffix:
<n>
<m>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
INITiate<n>:CONTinuous
Output unit
%
Hz
rad/deg (defined with
dB (Range Log or Range Linear %)
% (Range Linear dB)
.
Window
Marker
Result at the marker position.
INIT:CONT OFF
Switches to single measurement mode.
CALC:MARK2 ON
Switches marker 2.
INIT;*WAI
Starts a measurement and waits for the end.
CALC:MARK2:Y?
Outputs the measured value of marker 2.
Query only
See
"Marker Table"
See
"Marker Peak List"
.
Window
irrelevant
on page 295.
on page 279)
UNIT<n>:ANGLe
on page 22
on page 23
Analyzing Results