R&S FSW-K7 User Manual page 357

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
Suffix:
<n>
<k>
Parameters:
<Offset>
Manual operation:
CALCulate<n>:LIMit<k>:CONTrol:SHIFt <Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
<k>
Parameters:
<Distance>
Manual operation:
CALCulate<n>:LIMit<k>:CONTrol:SPACing <InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Suffix:
<n>
<k>
Parameters:
<InterpolMode>
Example:
CALCulate<n>:LIMit<k>:LOWer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Suffix:
<n>
<k>
User Manual 1173.9240.02 ─ 23
Remote Commands for Analog Demodulation Measurements
.
irrelevant
Limit line
Numeric value.
The unit depends on the scale of the x-axis.
*RST:
0
See
"X-Offset"
on page 156
.
irrelevant
Limit line
Numeric value.
The unit depends on the scale of the x-axis.
See
"Shift x"
on page 160
.
Window
Limit line
LINear | LOGarithmic
*RST:
LIN
CALC:LIM:CONT:SPAC LIN
.
irrelevant
Limit line
Analyzing Results
357