R&S FSW-K7 User Manual page 66

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
The following settings and functions are available to provide input via the optional Digi-
tal Baseband Interface in the applications that support it.
These settings are only available if the Digital Baseband Interface option is installed on
the R&S FSW.
For more information, see the R&S FSW I/Q Analyzer and I/Q Input User Manual.
Digital I/Q Input
Input Sample
Full Scale
Adjust Reference Level to Full Scale
Connected
DigIConf........................................................................................................................ 67
Digital I/Q Input State
Enables or disable the use of the "Digital IQ" input source for measurements.
"Digital IQ" is only available if the optional Digital Baseband Interface is installed.
Remote command:
INPut:SELect
Input Sample Rate
Defines the sample rate of the digital I/Q signal source. This sample rate must corre-
spond with the sample rate provided by the connected device, e.g. a generator.
If "Auto" is selected, the sample rate is adjusted automatically by the connected
device.
The allowed range is from 100 Hz to 10 GHz.
Remote command:
INPut:DIQ:SRATe
INPut:DIQ:SRATe:AUTO
User Manual 1173.9240.02 ─ 23
State.................................................................................................... 66
Rate........................................................................................................ 66
Level.............................................................................................................67
Instrument...................................................................................................67
on page 191
on page 215
on page 215
Input and Frontend Settings
Level...................................................................67
Configuration
66