R&S FSW-K7 User Manual page 41

Analog demodulation measurement option
Table of Contents

Advertisement

®
R&S
FSW-K7
If the DUT inserts a gain or an attenuation in the measurement, this effect can be
reflected in the result display on the R&S FSW. To reflect a power offset in the mea-
surement trace, change the
4.7.4.7
Coupling the Frequencies
As described in
ment results are very accurate as long as the same settings are used as for calibration.
Although approximate normalization is possible, it is important to consider the required
frequencies for calibration in advance. The frequencies and levels supported by the
connected signal generator are provided for reference with the interface configuration.
Two different methods are available to define the frequencies for calibration, that is to
couple the frequencies of the R&S FSW with those of the signal generator:
Manual coupling: a single frequency is defined
Automatic coupling: a series of frequencies is defined (one for each sweep
point), based on the current frequency at the RF input of the R&S FSW; the RF fre-
quency range covers the currently defined span of the R&S FSW (unless limited by
the range of the signal generator)
Automatic coupling
If automatic coupling is used, the output frequency of the generator (source frequency)
is calculated as follows:
Source
Equation 4-1: Output frequency of the generator
Where:
F
Generator
F
Analyzer
Numerator = multiplication factor for the current analyzer frequency
Denominator = division factor for the current analyzer frequency
F
Offset
converting measurements or harmonics measurements
The value range for the offset depends on the selected generator. The default setting
is 0 Hz. Offsets other than 0 Hz are indicated by the "FRQ" label in the channel bar
(see also
Swept frequency range
The F
the stop frequency defined in the "Frequency" settings of the R&S FSW. The resulting
output frequencies
in the "External Generator" > "Measurement Configuration" for reference.
User Manual 1173.9240.02 ─ 23
Reference
Chapter 4.7.4.5,
Numerator
Freq
RF
Offset
Denominato
r
= output frequency of the generator
= current frequency at the RF input of the R&S FSW
= frequency offset for the current analyzer frequency, for example for frequency-
Chapter 4.7.4.8, "Displayed Information and
values for the calibration sweep start with the start frequency and end with
Analyzer
(Result Frequency Start
Receiving Data Input and Providing Data Output
Value.
"Normalization",
on page 38, normalized measure-
Errors",
and
Result Frequency
Measurement Basics
on page 43).
Stop) are displayed
41

Advertisement

Table of Contents
loading

Table of Contents