Calibrant Reservoir Kit

Thermo Scientific
To replace the calibrant reservoir kit, which consists of the calibrant reservoir and its
cover
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components. If you just need to refill the calibrant reservoir and not replace it, see
Refilling the Calibrant
1. Shut down your system by following the steps in
on
page
139.
Note You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Unscrew the hydrogen safety screw.
3. Open the front door of the TSQ 9610 instrument.
4. Twist the calibrant reservoir cover counter-clockwise and remove it from the calibration
gas module.
Figure 427. Removing the Calibrant Reservoir Cover
Reservoir.
3
Advanced Troubleshooting
Replacing the Calibration Gas Components
"Shutting Down the TSQ 9610 System"
Calibration
Gas Module
Calibrant
Reservoir Cover