Sa Relative Level Accuracy Test - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Performance Tests
Individual Test Description

6. SA Relative Level Accuracy Test

Description
This test checks the relative level accuracy for the dpectrum monitor of the E5052B. The
mixed signals generated by two signal generators are used, and the carrier frequencies for
this test are 10 MHz, 1 GHz and 3 GHz.
Test Equipment
Signal Generator 1
Signal Generator 2 (or Sweeper)
Power Meter
Power Sensor
Step Attenuator
Power Splitter
N(m)-N(m) Adapter
Power Divider
N(m)-N(m) cable 61cm, 3 ea.
BNC cable, 122 cm, 2 ea.
24
Keysight E8257D with Opt. UNX, UNT, 1EA
without Opt. UNW, 1EH
Keysight E5070B/71B (use of the source)
Keysight E4419B
Keysight E9304A Opt. H18 or Keysight E9304A
Opt. H19
Keysight 8496H
Keysight 11667A
Keysight p/n 1250-1475
Keysight 11636A
Keysight N6314A (p/n 8120-8862)
Keysight p/n 8120-1840
Chapter 2

Advertisement

Table of Contents
loading