Setting Up Measurement Conditions And Display; Setting Up Measurement Conditions; Selecting Measurement Parameters - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Programming Manual
3

Setting Up Measurement Conditions and Display

This chapter describes how to set up the measurement conditions and display.
It also describes how to save/recall the instrument settings including the
measurement conditions.

Setting Up Measurement Conditions

Selecting measurement parameters

To select the measurement parameters, use the following commands.
You can select one of the following measurement parameter combinations
shown in
the resulting parameter combination is not among those in
other parameter is automatically changed to a proper parameter. For example,
when the primary parameter is Cp and you select Rs as the secondary
parameter, the primary parameter is automatically changed to Cs.
Table 3-1
Measurement parameter combinations
Primary parameter
Cp
Cs
Each parameter is described below.
Cp:
Cs:
D:
Q:
:CALCulate1:FORMat
:CALCulate2:FORMat
Table
3-1. If you select one parameter using the SCPI command and
Capacitance value measured using the parallel equivalent
circuit model
Capacitance value measured using the series equivalent
circuit model
Dissipation factor
Quality factor (inverse of D)
on page 188
on page 191
Secondary parameter
D, Q, G, Rp
D, Q, Rs
Table
3-1, the

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4981a/001E4981a/002

Table of Contents