X1, X2–Input to trigger circuit (plug-in unit) of Triac channels and coils of
power relays.
X3 pins 1 and 2–10 VAC input for 12 VDC unregulated power supply.
X3 pins 3 to 4–DC output (12 V) to electrical and condensation fans.
X2–Safety switches connection. Open=> no power to trigger circuit and
relay coils.
F, N (Phase, Neutral) Same as L1/L2–See the electrical diagram Triac/Relay
PCB.
Main power input terminals. The power source may have 2, 3, 4 wires
including ground.
1–57, printed circuit board terminals–Switched output to water solenoid
valve, heating elements, blower, circulation, replenishment pumps, main
motor, and step-down transformer.

2.3.2 The LUTH Power Board (LPB)

The Luth Power Board switches the necessary 2230V components on and off.
The LPB pcb includes:
Five long life zero cross Triac channels (only four are used)
Seven power relays with noise suppression circuits aross the contacts (only
six are used)
220V AC power supply to he prim. of the 24V AC transfo for the LMB02
and the crane hoist transfo
Voltage supply for the six safety switches. Four are for the top cover's wet
and dryer sections; two are the the front box (table lid) only in the latest
OLP version.
Two diodes for the half-wave rectified AC replenishment pumps.
2-6 Functional Analysis