Carriage Home Position Alignment Procedure - AGFA Avantra 36/44 Service Manual

Table of Contents

Advertisement

5. Enter the following Information:
Resolution:
Density Start (Enter Starting Density):
Resolution:
Density Start (Enter Starting Density):
Resolution:
Density Start (Enter Starting Density):
Resolution:
Density Start (Enter Starting Density):
6. Press ESC to exit out of this menu.
7. Select Image and then Test Files.
8. In the test file menu select EXP4RES.FIL and press ENTER.
The Avantra now images all four resolutions.
9. When imaging completes, select the Feed/Cut Icon on the Avantra's OCP.
10. Remove the take-up cassette and develop.
11. Using the densitometer, select the best 50% dot with a Dmax. between 3.8
to 4.2 for all four resolutions.
12. Enter the new density values in the Cassette Menu on the Avantra control
panel.
5.4
Carriage Home Position Alignment
Procedure
Use the Carriage Home Position Alignment Procedure to set the starting position
of the carriage assembly on the drum surface in the slow scan direction.
Position the carriage by entering a carriage home value in the AVDIAG program
under the DEC Machine Parameters. This value determines the distance between
the carriage limit sensor and the actual start of the image on the drum.
value is stored in EEPROM on the DEC.
Tools Required: 18" Scale, eye loupe
Steps:
1. Turn the system on and allow the system to boot the application software.
2. Connect the systems APIS cable to the PC Imager board in your PC.
Section I: Imagesetter
3600
Inc. 5
2400
Inc. 5
1800
Inc. 5
1200
Inc. 5
Alignment 5-5
The