2. Slide the control panel up and remove.
3. Disconnect the control panel cable and take it to the back of the processor.
4. Connect the control panel to the X400 with the remote cable.
WARNING:
handling.
CAUTION:
cessor is plugged in, or the CPU fuse will fail.
NOTE:
Potentiometer P102, N Min on the LUTH motor board, is not
used with infrared pulse transmission.
To Calibrate the LMB:
1. Turn off the processor from the control panel.
2. With the control panel off, enter the service program by pressing the fol-
lowing keys in sequence:
AUTO
DRYER SAVER
+
TIME
The display should read:
Service Code ___ .
3. Enter code 112 by pressing the following keys in sequence:
DEV
DEV
FIX
The display should read:
ACT 60 PRE 60
Where ACT = actual motor speed, and PRE = preset motor speed. Change
the preset value with the 15% and 85% keys, which decrease and increase
the value.
4-8 Maintenance
The PCB has high voltage terminals. Use caution when
Never connect or disconnect the cable when the pro-
Dev
Dev
Fix
Dev
Note: DEV
= 1
Fix
FIX
= 2
H
O
WAT
= 3
2