4. When the motor is turning in the wrong direction, switch the motor off
immediately by pressing the 50% button on the control panel. Exchange
the motor wires M1 and M2.
5. Set the speed to 20 seconds and adjust P203 (N-max.) until 20 ± 1 seconds
is obtained. To increase the speed, turn P203 clockwise.
6. Set the speed to 50 seconds and adjust P219 (N-min.) until 50 ± 2 seconds is
obtained. To increase the speed, turn P219 clockwise. Repeat 20 and 50
seconds and readjust, if necessary.
7. Switch the processor to normal operation and make a functional checkout.
NOTE:
Remember to reprogram preset values in the control panel.
The motor overload message in the display (only visual in normal mode,
not in service mode) appears in this EPROM version as:
Disconnect Mains + Motor Off. The following table shows jumper
settings for current limiter cut-out levels:
Jumper
Position
10
9
8
7
6
5
4
3
2
1
Connections:
X1-1/2/4
input optical interruptor
X2
not used
X3-1/2
tacho feedback connected to X503 -B/14 LMB
X4A/B
not used
X102-1
M+ to the drive motor
-2
M- to the drive motor
-3/4
not used
-5/6
24VAC input from transformer
Current
Limit
Machine Type
0.5A
1.0A
1.5A
2.0A
2.5A
Rapiline 43-3, 51-3. 66-3, 72-3
3.0A
RL 72 OLP for Dolev, RL 51-3 OLP
Accuset, Ecorap 51 & 72
3.5A
Avantra 44 P and OLP
4.0A
Avantra 44
4.5A
5.0A
Section IV: On-line Processor
Maintenance 4-9