Operational Inspection - AGFA Avantra 36/44 Service Manual

Table of Contents

Advertisement

Avantra 36/44 Service Manual
4. Open the right side cover and remove the rear access panel by pushing up
on the three white buttons. Remove the access panel.
5. Remove the two screws securing the back supply light shield located
behind the supply compartment. Pull the light shield up between the sup-
ply compartment and the bridge.
6. Lift the panel covering the supply punch and the punch container.
7. Vacuum out punch chaffs from the punch container pan from the right.
8. Re-install the panel and covers.
6.2.1.5
General Inspection/Cleaning for Customer
The general inspection/cleaning should be performed by the customer.
To Inspect/Clean:
1. Vacuum the supply cassettes before installing a new roll of media.
2. Vacuum the black intake fan filter (P/N 66602-001) every month.
3. Replace the exhaust fan filter (P/N C00305-001 Orange/White/Black) every
three months.
4. Clean the punch remnants containers every three months.

6.2.2 Operational Inspection

This inspection insures the quality and proper alignment of the system.
To Inspect:
1. Turn the system on and monitor the boot-up sequence.
2. Insure that the system successfully completes BCC's and loads the applica-
tion firmware with no errors.
3. Using the Avantra Diagnostic (AVDIAG) program on the PC, run all select-
able tests. Insure that all tests run successfully.
4. Check the error log, noting any errors. Discuss the results with the opera-
tor. Before exiting, clear the error log.
5. Using the PC Imager program, run an exposure array @ 2400dpi.
6. Check the density and enter the appropriate value (if different) in the
Parameters submenu.
7. Run the following test files:
50% @ 2400
100/85% @ 2400
Scale/Fid @ 2400
1x1 Hor. Line @ 2400
Examine the output copy and resolve any problems.
6-4 Maintenance