28.
Check internal wiring between X501 plug and input sensors.
29.
Check that circulation pumps are running during heating.
30.
Verify operation of electronic system with service code 111.
31.
Verify operation of electronic system with service code 121.
32.
Verify operation of electronic system with service code 123.
33.
Verify operation of electronic system with service code 131.
34.
Verify operation of electronic system with service code 132.
35.
Verify operation of electronic system with service code 211.
36.
Verify operation of electronic system with service code 212.
37.
Verify operation of electronic system with service code 213.
38.
Verify operation of electronic system with service code 221.
39.
Verify operation of electronic system with service code 231.
40.
Verify operation of electronic system with service code 232.
41.
Verify operation of electronic system with service code 233.
42.
Verify operation of electronic system with service code 311.
43.
Verify operation of electronic system with service code 312.
44.
Verify operation of electronic system with service code 331.
45.
Blower capacitor defective. Replace.
46.
Verify operation of electronic system with service code 333.
47.
Built-in thermo fuse in heating element is open.
48.
Incorrect trigger signal to power triac. Replace trigger board.
49.
Time not adjusted on ATB. Insert correct time.
50.
Clock not working due to battery failure. Check readout over a period of time. Check
battery. Connect a DVM across the terminals. Normal reading should be 3.0 V. If volt-
age is lower than 2.5 V., battery should be replaced. Note that normal life span is
approximately 10 years.
51.
Check quality of each connection between developer heating element and fixer heat-
ing element.
52.
Check replenishment empty sensor wiring and plugs for continuity; i.e., no open or
short circuits. Check that the sensors are not set so low as to come in contact with the
bottom of the replenishment tank.
53.
Replenishment tubes not in proper position in replenishment container, or replenish-
ment container is empty.
54.
Recalibrate pumps using MAN REPL key. See Service Program.
55.
Check length of brushes.
56.
Adjust LMB as described in the LMB Calibration Procedure in Section 4.3 on page 4-
6.
57.
Check flow control at water solenoid valve inlet.
Table 4-4 Repair steps (continued from previous page).
4-22 Maintenance

Advertisement

Table of Contents
loading

This manual is also suitable for:

Avantra 36s/44s

Table of Contents