Refilling The Calibration Standard Vials With Fresh Standard - Thermo Scientific Orion AQUAfast AQ3010 User Manual

Turbidity meter
Table of Contents

Advertisement

Refilling the
Calibration
Standard
Vials with
Fresh
Standard
10
1 . Pour the old standard out of the vial .
2 . Wash the vial with laboratory glassware detergent and water . Rinse
the vial with turbidity-free water to remove all laboratory detergent
from the vial .
3 . Rinse the vial with about 5 mL of the standard from the plastic bottle
with the matching NTU value .
4 . Cap and swirl the vial so the inside of the vial is fully rinsed with the
fresh standard .
5 . Pour the standard out of the vial .
6 . Pour the fresh standard into the vial up to the mark on the vial and
cap the vial .
7 . Perform the procedure in the Calibration Check section with the vial
containing the fresh standard .
If the fresh standard does not read with the 10% range of the nominal
value, or to the tolerance of your method, follow the steps in the
Calibration Procedure section .