SCPI.CALCulate(Ch).FSIMulator.BALun.PARameter.TNAMe.SSBALanced.DE
Fine(Name)
Object type
Property (Read-Write)
Syntax
SCPI.CALCulate(Ch).FSIMulator.BALun.PARameter.TNAMe.SSBalanced.DEF
ine(Name) = Param
Param =
SCPI.CALCulate(Ch).FSIMulator.BALun.PARameter.TNAMe.SSBalanced.DEF
ine(Name)
Description
This command sets/gets the measurement parameter when the balance
device type is "unbalance-unbalance-balance", for the selected trace of the
selected channel (Ch). The trace can be selected by the trace name once
its defined earlier with SCPI.CALCulate(Ch).PARameter(Tr).TNAMe.DATA
command. This function is available in the revision A.9.60 and above.
Variable
Parameter
Description
Data type
Parameter
Description
Data type
Range
1598
Name
Trace Name
Character string type (String)
Param
Measurement parameter
Character string type (String)
Select from the following.
"SSS11": Specifies Sss11.
"SSS21": Specifies Sss21.
"SSS12": Specifies Sss12.
"SSS22": Specifies Sss22.
"SDS31": Specifies Sds31.

Hide quick links: