Agilent Technologies E5071C Manual page 1376

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

SCPI.CALCulate(Ch).FSIMulator.BALun.PARameter(Tr).BALanced.DEFine
Object type
Property (Read-Write)
Syntax
SCPI.CALCulate(Ch).FSIMulator.BALun.PARameter(Tr).BALanced.DEFine =
Param
Param =
SCPI.CALCulate(Ch).FSIMulator.BALun.PARameter(Tr).BALanced.DEFine
Description
This command sets/gets the measurement parameter when the balance
device type is "balance", for the selected trace (Tr) of the selected channel
(Ch).
This command is available at the firmware revision A.9.50 and above.
Variable
Parameter
Description
Data type
Range
Preset
value
Examples
Dim BalPara As String
SCPI.CALCulate(1).FSIMulator.BALun.DEVice = "bal"
SCPI.CALCulate(1).FSIMulator.BALun.PARameter(1).BALanced.DEFine = "sdd11"
BalPara = SCPI.CALCulate(1).FSIMulator.BALun.PARameter(1).BALanced.DEFine
Related objects
SCPI.CALCulate(Ch).FSIMulator.BALun.DEVice
Equivalent key
> Fixture Simulator|[Meas] > Sdd11|Scd11|Sdc11|Scc11
Analysis
Equivalent SCPI command
Param
Measurement parameter
Character string type (String)
Select from the following.
"SDD11": Specifies Sdd11.
"SCD11": Specifies Scd11.
"SDC11": Specifies Sdc11.
"SCC11": Specifies Scc11.
"SDD11"
Programming
1603